Kuwait University
COLLEGE OF ENGINEERING & PETROLEUM

Low Level Sulfur Analyzer

Principle:
The principle for the operation of sulfur analysis begins with the complete, high temperature oxidation of the entire sample matrix. The sample is combusted with oxygen at a temperature of 1050 ºC. The combustion gases are routed through a membrane drying system to remove all water and then to the sulfur detector module for quantization. The SO2 is exposed to ultraviolet radiation of a specific wavelength. This radiation is re-released in the form of sulfur fluorescence. The fluorescence is detected by a photomultiplier tube (PMT) and it is proportional to the amount of sulfur in the original sample.

Applications:

  • Petrochemicals.
  • Ultra low sulfur content fuel.
  • Pharmaceutical Industry.
  • Green Manufacturing.


Specification and standards:

  • ASTM D7183 - 07e1
  • Sulfur detection range 1…25 ppm


Manufacturer and Model No.
Manufacturer: Antek by PAC instrumentation
Model No.: Multitek


















Location: Bldg No. 8Kh, 208
Contact: Ext. 3621